Laboratory preparation of aqueous Zn-Zn symmetric cell deposition and dissolution test

Latest updated: Aug 05, 2024
Video recording version: 8.0.0.513, compatible with BTS 8.0 software.

Today, we will demonstrate the test step settings for the cycle performance of Zn-Zn symmetric batteries.

The current will be set according to the current density of 1 mA/cm2. Taking an electrode with a diameter of 1 cm as an example, the current will be 0.785 mA based on the current density and the electrode area.

Open the BTS software, right-click on the channel where the battery is being held, and click “Single start.” This will open the work step settings interface.

The rest time is set to 10 minute, the charge and discharge time is set to 1 hour, and the current is set to 0.785 mA.

Enter the cycle count. Without knowing exactly how many cycles the symmetric battery can be cycled, it is recommended to set a higher number of cycles. Once the test voltage is disrupted, stop the test. Save the step and start the test.

The above working step parameters are set up for reference only, specifically, please set them according to the actual experimental conditions. Thank you for watching.

This operation video tutorial is applicable to the following products:
● CT-4000 Battery Testing System Series (CT-4008Q-5V100mA, CT-4008Q-5V6A-S1, CT-4008Q-5V6A-164, CT-4008Q-5V12A-S1, CT-4008Q-5V12A-164, etc.)
● CE-6000 Battery Testing System Series (Cell, Module, Pack, etc.)
● CT/CTE-8000 Battery Testing System Series (CT-8002Q-5V100mA, CTE-8008-5V200A, etc.)
● All-in-One Battery Testing System Series (WHW-25L-S-16CH, WIHW-200-160CH-B, WHW-100L-2-160CH, WHW-500L-5V6A80CH-380V, WGDW-380L-2-40BFC-5V600A8CH, etc.)